Scientific Library of Tomsk State University

   E-catalog        

Image from Google Jackets
Normal view MARC view

Model of optical phantoms thermal response upon irradiation with 975 nm dermatological laser M. S. Wróbel, A. N. Bashkatov, A. N. Yakunin [et al.]

Contributor(s): Bashkatov, Alexey N | Yakunin, Alexander N | Avetisyan, Yuri A | Genina, Elina A | Galla, S | Sękowska, A | Truchanowicz, D | Cenian, A | Jędrzejewska-Szczerska, Małgorzata | Wróbel, Maciej S | Tuchin, Valery VMaterial type: ArticleArticleSubject(s): численное моделирование | оптические фантомы | термография | лазеры | медицинаGenre/Form: статьи в журналах Online resources: Click here to access online In: Proceedings of SPIE Vol. 10716 : Saratov fall meeting 2017 : Optical technologies in biophysics and medicine XIX, 26-30 September 2017, Saratov, Russian Federation. P. 1071604-1-1071604-6Abstract: We have developed a numerical model describing the optical and thermal behavior of optical tissue phantoms upon laser irradiation. According to our previous studies, the phantoms can be used as substitute of real skin from the optical, as well as thermal point of view. However, the thermal parameters are not entirely similar to those of real tissues thus there is a need to develop mathematical model, describing the thermal and optical response of such materials. This will facilitate the correction factors, which would be invaluable in translation between measurements on skin phantom to real tissues, and gave a good representation of a real case application. Here, we present the model dependent on the data of our optical phantoms fabricated and measured in our previous preliminary study. The ambiguity between the modeling and the thermal measurements depend on lack of accurate knowledge of material's thermal properties and some exact parameters of the laser beam. Those parameters were varied in the simulation, to provide an overview of possible parameters' ranges and the magnitude of thermal response.
Tags from this library: No tags from this library for this title. Log in to add tags.
No physical items for this record

Библиогр.: 24 назв.

We have developed a numerical model describing the optical and thermal behavior of optical tissue phantoms upon laser irradiation. According to our previous studies, the phantoms can be used as substitute of real skin from the optical, as well as thermal point of view. However, the thermal parameters are not entirely similar to those of real tissues thus there is a need to develop mathematical model, describing the thermal and optical response of such materials. This will facilitate the correction factors, which would be invaluable in translation between measurements on skin phantom to real tissues, and gave a good representation of a real case application.
Here, we present the model dependent on the data of our optical phantoms fabricated and measured in our previous preliminary study. The ambiguity between the modeling and the thermal measurements depend on lack of accurate knowledge of material's thermal properties and some exact parameters of the laser beam. Those parameters were varied in the simulation, to provide an overview of possible parameters' ranges and the magnitude of thermal response.

There are no comments on this title.